R&S
UD1065
Risk of injury and instrument damage if stacking instruments
A stack of instruments may tilt over and cause injury. Furthermore, the
instruments at the bottom of the stack may be damaged due to the load
imposed by the instruments on top.
Observe the following instructions when stacking instruments:
● Never stack more than three instruments. If you need to stack more
than three instruments, install them in a rack.
● The overall load imposed on the lowest instrument must not exceed
500 N.
● All instruments must have the same dimensions (width and length).
● If the instruments have foldable feet, fold them in completely.
Mounting the R&S UD1065 in a rack
The R&S UD1065 can be installed in a rack using a rack adapter kit (order no.
see data sheet). The installation instructions are part of the adapter kit.
Risk of instrument damage in a rack
An insufficient airflow can cause the instrument to overheat, which may dis-
turb the operation and even cause damage.
Make sure that all fan openings are unobstructed, that the airflow perfora-
tions are unimpeded, and that the minimum distance from the wall is 10 cm.
Getting Started 1178.4176.02 ─ 02
Preparing for Use
Placing or Mounting the Instrument
11