CHANGING THE FILTER
Never operate the unit without a filter.
Replace the filter after every ten uses, or sooner if it turns gray, appears contaminated
is blocked or becomes wet.
Use only Veridian Healthcare filters; additional filters are available for order using the
included order form, calling Customer Care at 866-326-1313 or contacting your local
dealer.
Do not substitute cotton or any other materials for the filters.
Do not attempt to clean the filters or reuse a filter.
To replace the filter, follow these steps:
1. Change the filter only when the unit is off,
unplugged and completely cool.
2. Gently pull the filter cap out of the
compressor.
3. Remove the filter from the filter cap.
4. Place a new filter in the filter cap.
5. Return the filter cap to its place, ensuring
that it is firmly in place.
14 • ENGLISH