5. SPECIFICATIONS
5.1
Electrical
Infrared
●
Temperature Scale: Celsius (°C) or Fahrenheit (°F) user-selectable
●
Temperature Range: -4°F to 1022°F (-20°C to 550°C)
●
Display Resolution: 0.1°C/°F or 1°C/°F
●
Accuracy: ±2% of reading or ±6°F/3°C, whichever is greater at 64.4
to 82.4°F (at 18 to 28°C) ambient operating temperature.
●
Temperature Coefficient: Changes in accuracy operating temperature
above 82.4°F/28°C or below 64.4°F/18°C: ±0.2% of reading or ±0.36°F/
0.2°C, whichever is greater.
●
Response Time: 1 second
●
Laser: Red, <1mW (670nm) Class II, 2 - 50ft Range.
●
Spectral Response: 6 to 14µm nominal
●
Emissivity: Pre-set at 0.95, user selectable from 0.10 to 1.00
●
Detection Element: Thermopile
●
Optical Lens: Fresnel Lens
●
Field of View (FOV) ratio: 10:1 (Distance: Diameter)
K Type
●
Temperature Scale: Celsius (°C) or Fahrenheit (°F) - User-selectable
●
Temperature range of the instrument in function of the K type
thermocouple chosen (see § 6) : - 40 to +1350°C (-40 to 2000°F)
-40 to 200°C (-40 to 392°F) with the K type thermocouple supplied
as standard.
●
Resolution: 0.1°C/°F or 1°C/°F
●
Accuracy (meter):
-40°C to 1350°C: ±(0.1% rdg ± 1°C) on -40°F
to 2000°F: ±(0.1% rdg ± 2°F) (Accuracy is specified for operating
temperatures over the range of 18°C to 28°C, for 1year, excluding
the sensor)
●
Temperature Coefficient: 0.1 times the applicable accuracy
specification per °C from 0°C to 18°C and 28°C to 50°C
●
Input Protection: 24VDC or 24Vrms maximum input voltage on any
combination of inputs.
●
Sample Rate: 2.5 times per second.
●
Input Connector: Standard miniature thermocouple connectors (flat
blades spaced 7.9mm, center to center).
●
Temperature Response: Temperature indication follows Reference
Temperature/Voltage Tables N.I.S.T. Monograph 175 Revised to ITS
-90 for K-type thermocouples.
5.2 General specifications
●
Display: 3½ digit liquid crystal display (LCD) with maximum reading
of 1999
●
Over-Range: (OL) or (-OL) is displayed
●
Low battery indication: " symbole batterie " is displayed when the
battery voltage drops below the operating level
●
Sample rate: 2.5 times per second, nominal.
●
Environment:
Operating:
32°F to 122°F (0°C to 50°C) at < 80% relative humidity
Storage:
-4°F to 140°F (-20°C to 60°C), 0 to 80% RH with
battery removed from meter
●
Auto power OFF: 15 seconds approx.
●
Altitude: 2000m max.
●
Input Protection: 24VDC or 24rms maximum input voltage on any
combination of inputs.
19