Protection Mode Error Codes
Abnormal Code
E01
E02
E03
E04
E05
E06
E07
E08
E09
E10
E11
E12
E13
The height difference between
the two legs of the controller
exceeds the set value and is
not synchronized.
H01
Abnormal Description
M1 Over Current Protection
M2 Over Current Protection
M3 Over Current Protection
M4 Over Current Protection
M5 Over Current Protection
M6 Over Current Protection
M1 Hall Fault
M2 Hall Fault
M3 Hall Fault
M4 Hall Fault
M5 Hall Fault
M6 Hall Fault
Communication interruption
between controller A and B.
Overheat Protection
(Work Mode Protection)
Check if the load is too heavy or if there
is any object blocking the rise and
lowering motion of the desk.
If the issues from above mentioned are
worked out, long-press the down button
to reset, and after the reset is completed,
continue using it.
Check if the legs are tightened, after
eliminating the anomaly, long-press the
down button to reset, and after the reset
is completed, continue using it.
Check if the connector between the A and
B controllers is disconnected. After
eliminating the anomaly, long-press the
down button to reset, and after the reset
is completed, continue using it.
Check if the legs are tightened, after
eliminating the anomaly, long-press the
down button to reset, and after the reset
is completed, continue using it.
The controller needs to rest for a while.
After the temperature drops, it can
continue to be used.
12
Solution