Descargar Imprimir esta página

Toolland DVM852 Manual Del Usuario página 9

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 36
ranging mode ................................................................ manual
ranges ................................................................................. 19
data hold ............................................................................ yes
backlight ............................................................................. yes
auto power-off ...................................................................... no
dimensions ..................................................... 68 x 138 x 30 mm
weight (with battery) ..................................................... ± 170 g
storage temperature ............................................ -10 °C to 50 °C
test lead probe (incl.) ..................... CAT III 600 V, 10 A; L = 80 cm
9.1 DC VOLTAGE
range
200 mV
2000 mV
20 V
200 V
600 V
Overload protection: 250 V rms for 200 mV range; 600 V DC or AC rms for
all other ranges
9.2 AC VOLTAGE
range
200 V
600 V
Average sensing, calibrated to rms of sine wave
Frequency range: 40-400 Hz
Overload protection: 600 V DC or AC rms
V. 01 – 29/03/2016
All manuals and user guides at all-guidesbox.com
DVM852
Do not measure circuits that may contain voltages
resolution
0.1 mV
1 mV
10 mV
100 mV
1 V
Do not measure circuits that may contain voltages
resolution
100 mV
1 V
> 600 V
accuracy
± (0.5 % rdg + 2 digits)
± (0.8 % rdg + 2 digits)
> 600 V
accuracy
± (1.2 % rdg + 10 digits)
9
©Velleman nv

Publicidad