Descargar Imprimir esta página

Apex Domus 2+ Manual De Instrucciones página 17

Ocultar thumbs Ver también para Domus 2+:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 20
Domus 2+ / English
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make
sure it is used in such an environment.
Immunity Test Levels
Professional
Basic EMC standard
healthcare
facility environment
Electrostatic
±8kV contact
Discharge (ESD)
±15kV air
IEC61000-4-2
Electrical fast
±2kV for power supply line
transient/ burst
±1kV for input/output line
IEC61000-4-4
± 1 kV line(s) to
Surge
line(s)
IEC61000-4-5
± 2 kV line(s) to earth
Voltage dips, short
Voltage Dips:
interruptions and
i) 100% reduction for 0.5 period,
voltage variations
ii) 100% reduction for 1 period,
on power supply
iii) 30% reduction for 25/30 period,
input lines
Voltage Interruptions:
IEC61000-4-11
100% reduction for 250/300 period
Power frequency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz – 80 MHz
Conducted RF
6 Vrms in ISM bands
IEC 61000-4-6
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
3 V/m 80 MHz to
2.7 GHz
80 % AM at 1 kHz
Radiated RF EM
Fields
385-6000 MHz,
IEC61000-4-3
9-28V/m, 80%
AM(1kHz) pulse mode
and other modulation
Compliance
HOME HEALTHCARE
Levels
ENVIRONMENT
±8kV
contact
±15kV air
±2kV for
power supply
line
±1kV for
input/outpu
t line
± 1 kV line(s) to
± 1 kV
line(s)
line(s) to
line(s)
230V
30 A/m
30 A/m
3 Vrms
0,15 MHz – 80 MHz
6 Vrms in ISM and
amateur radio
6Vrms
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
10 V/m 80 MHz to
2,7 GHz
80 % AM at 1 kHz
10V/m
385-6000 MHz,
9-28V/m, 80%
AM(1kHz) pulse
mode and other
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of
atypical commercial or hospital
environment
Mains power quality should be that of
atypical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of this device
requires continued operation during
power mains interruptions, it is
recommended that the device be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Portable and mobile RF
communications equipment should be
used no closer to any part of this
device, including cables, than there
commended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d =
150kHz to 80MHz
P
=
80MHz to 800MHz
d
0
6 ,
P
=
800 MHz to 2.7GHz
d
1
2 ,
P
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
14