4.3.7. Temperature measurement with K probe (HT62)
Before attempting any temperature measurement, cut off power supply from
the circuit to be measured and make sure that all capacitors are discharged,
if present.
Fig. 8: Use of the instrument for Temperature measurement
1. Select position Temp°C°F
2. Press the MODE key until the symbol
3. Insert the provided adapter into input terminals VHz%CAPTemp
COM (polarity -) (see Fig. 8)
4. Connect the provided K-type wire probe or the optional K-type thermocouple (see § ) to
the instrument by means of the adapter, respecting the positive and negative polarity
on it. The display shows the value of temperature.
5. The message
measurable value.
6. To use the HOLD function, see § 4.2.
O.L.
indicates that the value of temperature exceeds the maximum
"
"
CAUTION
°C
or
°F
is displayed.
"
"
"
"
EN - 14
HT61-HT62
(polarity +) and