ELECTROMAGNETIC COMPATIBILITY
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE
RF COMMUNICATIONS EQUIPMENT AND THE EQUIPMENT OR SYSTEM
FOR EQUIPMENT OR SYSTEM THAT NOT LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and
the 11-50DP Pulse Oximeter
The Pulse Oximeter is intended for use in the electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Pulse Oximeter can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Pulse Oximeter as recommended below, according
to the maximum output power of the communications equipment.
Rated maximum
150KHz to
output power of
80MHz
transmitter
(W)
0.12
0.01
0.37
0.1
1.17
1
3.69
10
11.67
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distanced in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter (m)
80MHz to
800MHz
0.12
0.37
1.17
3.69
11.67
800MHz to
2.5GHz
0.23
0.74
2.33
7.38
23.33