Emergency stop
To activate emergency stop funktion: Pull the red emergency stop strap firmly, fig 1, and then
release.
To reset: Push the plastic button that holds the red strap in place upward until it clicks,
see fig 2.
When the emergency stop funtion is activated, a red indication is visible, fig 1, above
emergency stop strap and its white plastic button.
If the emergency stop strap is held down constantly, electrical emergency lowering will be
activated. When the strap is released, the lift will stop.
Electrical emergency lowering
If the hand control is not functioning, electrical emergency lowering can be activated by
holding the emergency stop strap down, fig 1. Lower the user to a safe height and location.
Electrical emergency lowering may only be used if the hand control is
malfunctioning.
The electrical emergency lowering device feeds out the lift strap for as long as the emergency
stop strap is held down. Ensure that the lowering motion is activated. Otherwise, electrical
emergency lowering must be stopped. Always ensure that emergency lowering is done in an
appropriate location.
To reset: Press the plastic button that holds the red strap in place upward until it clicks, see fig
2.
Manual emergency lowering/raising
Remove the plastic cover on the same side as the emergency stop strap, fig 3. On the inside
of the cover there is an Allen key. Place the supplied Allen key in the hole as indicated, fig 4.
Turn it counter-clockwise to lower and clockwise to raise the sling bar. Always ensure that
emergency lowering is done in an appropriate location.
NOTE! Lowering is very slow.
Re-setting manual emergency lowering/raising: Contact an authorized Handicare service techni-
cian. Do not use the lift unit.
SystemRoMedic
M A N U A L
TM
1
2
3
4
Allen key
15