Screen 1: Setting the temperature
Your Grant digital ultrasonic bath is fitted with a thermostat and heaters to ensure the
temperature of the cleaning liquid is maintained at the set temperature.
Use the UP and DOWN keys to accurately set the desired temperature of the cleaning
liquid between ambient and 70°C.
During normal use, the water heaters in your Grant digital ultrasonic bath will switch on
if
the
cleaning
[As a safety feature, your Grant digital ultrasonic bath will only turn on when the bath is
full of liquid.]
Please note that during normal operation, ultrasonic energy will heat the cleaning liquid
by up to 15°C per hour.
Your Grant digital ultrasonic bath cannot cool the cleaning liquid. If the liquid
temperature becomes too hot, you must either let it cool down, or refill the bath with
cooler water.
When the correct temperature is selected, use the MENU key to move to the next screen.
SAFETY NOTICE: When the temperature is set above 54°C, the case of the Grant
digital ultrasonic bath will become hot and care should be taken when touching the case.
A warning label is included in the literature pack that may be applied to the
unit if deemed necessary following a risk assessment
(see ISO 13732-1:2006).
xub Series
Operating Manual
liquid
temperature
is
lower
than
9
the
set
temperature.