Screen 6: Liquid level sensor
The Grant digital ultrasonic bath is fitted with a liquid level sensor to ensure it is not
under-filled prior to or during the cycle. This ensures that the optimum cleaning
process can occur and potential damage to the tank is minimised.
The level sensor is a resistive device, so will not detect liquids with very low
conductivity such as pure deionised (DI) water.
The level sensor can be disabled for specialist applications where deionised (DI)
water is used.
Use the ON and OFF keys to set the Low Level function. When the desired option is
selected, use the MENU key to move to the next screen.
Screen 6a: Liquid Level Sensor
When the OFF button is pressed and the  key is selected to continue, a message
will appear to check that the operator has filled the tank up to the level indicated (see
page 5).
When the level sensor is disabled the liquid set temperature is automatically reduced
EN
to 5°C. The set temperature may be increased by entering the MENU system (see
page 9).
Screen 6b: Liquid Level Sensor
If 'NO' is selected an instruction will appear requesting the tank be filled before use.
xub Series
19
29666 V5
Operating Manual