HBS x-touch 300
4 Specifications
Mains connection
Supply voltage
Nominal frequency
Output power in monopolar operation
Pure cut
Blend cut
Polypectomy mode
Cutting in water environment (TUR)
Soft coagulation
Forced coagulation
Spray coagulation
Output power in bipolar operation
Bipolar Coagulation
Bipolar Blend Cut
Bipolar Cutting in water environment (TUR)
AUTO STOP
Safety measures
Type
Protection class
Equipotential connection
Safety
Electromagnetic compatibility
Safety of high-frequency surgical units
Miscellaneous
Operating temperature
Humidity during operation
Air pressure during operation
Storage temperature
Storage humidity
Air pressure during storage
Dimensions (L/W/H)
Weight
100-260 Volt, 50 - 60 Hz.
500 kHz
300 W at 500 Ω
250 W at 500 Ω
100 W (peak: 250W) at 250 Ω
300 W at 500 Ω
150 W at 250 Ω
150 W at 500 Ω
80 W at 1500 Ω
100 W at 100 Ω
100 W at 100 Ω
120 W at 100 Ω
Yes
CF
I
Yes
In accordance with DIN EN ISO 60601-1
In accordance with DIN EN ISO 60601-1-2
In accordance with DIN EN ISO 60601-2-2
+10° C to +40°C
30-75%
700-1060hPa
-10° C to +60°C
10-75%
500-1060hPa
424 x 397 x 166 mm
6,9 kg