MANUAL DE INSTRUCCIONES
J.P. SELECTA, sau
CONC./ESTANDAR
SELECT
ENTER
CONC./ESTANDAR
CONC./ESTANDAR
PONER BLANCO
VALOR STD: 0.000
... Insert cell with blank
and press ENTER
...
ENTER
J.P. SELECTA, sa
LECTURA CONTINUA
ENTER
J.P. SELECTA, sa
CORRECCION ABS
J.P. SELECTA
, s.a.u.
CODIGO 80269 REV C
ENTER
CONC./ESTANDAR
... Input the standard
... Insert cell with stan-
value and press
ENTER ...
LECTURA CONTINUA
PONER BLANCO
... Insert cell with blank
and press ENTER
LECTURA CONTINUA
FACTOR: 1.00
... Enter the factor value and press
ENTER ...
(Do not insert the decimal point)
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - http://www.jpselecta.es
4.4
MEASURING CONCENTRATION BY
STANDARDS
Calculates the sample concentration measuring the absorbance and
applying a pre-introduced concentration factor. Manually select the
appropriate wavelength filter. To get good precision, use the same type
of flask with distilled water to made the blank.
ENTER
J.P. SELECTA, sau
PONER STD
PONER MUESTRA
... Insert cell with sample
dard and press
ENTER ...
4.5 MEASURING ABSORBANCE IN CONTINUOUS
READING
Continuous reading of the absorbance value of a sample every 4 times
per second. Manually select the appropiate filter.
ENTER
... Insert cell with a sample. Displays
...
4.6 ENTERING A CORRECTION FACTOR
This menu allows the user to enter a correction factor that is applied to
the absorbance reading. The default value is 1.00. The value needs to
be between 0.8 to 1.2.
Entering this factor is specially useful for repetitive samples whereby an
adjustement is required by a standard.
ENTER
... Stores the entered value (Range 0.8...1.2)
ESC
... Rejects the value entered.
05/2023
ENTER
CONC./ESTANDAR
C= 0.000 STD=0.000
... Display the concen-
and press ENTER
...
LECTURA CONTINUA
ABS= 0.000
the absorbance reading ...
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Pag.: 15
CLEAR
ENTER
ZERO
ESC
... Return to main
tration reading ...
menu ...
CLEAR
CERO
ESC
... Return to main
menu ...