Tare Function P3962
In this mode the positive deviation of the measured value from the
tared value (max. 0% of the measuring range) is constantly indica-
ted by the main display. The untared pressure value is always shown
on the bar graph display. Thus the digital manometer's actual wor-
king load can be read off even if the tare function is active. When the
tare function is active, a function indicator
There are  possible ways of activating this mode:
1) Select the tare function and set the value to be tared using
the keys
backed up as tare. The main display and the auxiliary display then
show the value "0000" in the unpressurised state. Example: If a
value to be tared is set as  bar, the main display indicates 1 bar
when the actual pressure is 3 bar.
) Load the digital manometer with the pressure to be tared.
The current measured value appears in the main display
field of the main display. Select the tare function and then
operate both keys simultaneously
value will be backed up as tare. The main display and the auxiliary
display are set to (0000).
Setting the tare value to 0000 resets the tare function in both
models.