Descargar Imprimir esta página

Mopedia Trendelenburg MP393 Manual De Instrucciones página 27

Publicidad

Idiomas disponibles

Idiomas disponibles

Table 3
Manufacturer's guide and declaration - electromagnetic immunity
The device is intended to be used in the following electromagnetic conditions. The client or user
must ensure that it is used in these conditions. Make sure it is used in these conditions.
Immunity tests
conducted RF
IEC 61000-4-6
radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, apply the separation distance for the higher frequency band.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
The intensity of the field for fixed transmitters, such as base stations for radio-telephones (cellulars and cordless)
a
and terrestrial mobile radios, amateur radio equipment, AM and FM radio transmitters and TV transmitters may not
be known precisely in theory. To establish an electromagnetic environment caused by fixed RF transmitters, one must run
an on-site electromagnetic inspection. If the field intensity measured in the place where the DEVICE is used exceeds the
applicable RF level of conformity indicated above, normal functioning of the DEVICE must be monitored. If abnormalities
are observed, supplementary measures may be necessary, such as re-orientation or repositioning of the DEVICE.
The field intensity over the frequency range from 150 kHz to 80 MHz should be less than 3 V/m.
b
Test level IEC 60601
3 V eff.
From 150 KHz to 80 MHz
3 V/m
From 80 MHz to 2.5 GHz
Conformity level
environment – guide
3 V eff.
Portable and mobile RF
communications devices
should not be used near
any part of the device,
including the cables,
unless the recommended
separation distances are
observed, calculated from the
equation applicable to the
transmitter frequency.
Recommended distances
From 80 MHz to 800 Mhz
from 800 MHz to 2.5 GHz
during charging operations
3 V/m
where "P" indicates the
maximum rated output
of the transmitter in Watt (W)
according to the transmitter
manufacturer and "d" is the
recommended separation
distance in metres (m).
The intensity of the field
generated by fixed RF
transmitters, as determined
in an on-site electromagnetic
inspection, might be below
the level of conformity in each
frequency band. Interference
may be experienced
in the proximity of equipment
marked with this symbol:
11
Electromagnetic
d = 1.2√P
d = 1.2√P
d = 2.3√P
d = 7.7 m