The rotor and accessories are rated for the maximum rotation frequency shown in the table below.
Tube Capacity
Dimensions (H x W x D)
Weight
Sound Level
Environmental Range
Voltage
Frequency
Power Requirement
Centrifuge Motor
Cycle Time (in Minutes)
Max Speed
Max Carrier Load
C
ALCULATING THE
The I.F.U.s of tube manufacturers recommend cycles at a minimum G-Force, which can be calculated if you know
the RPM and the radius. Use the formula below or go to www.druckerdiagnostics.com/g-force-calculator/.
In Centimeters:
RCF or G-force = 0.00001118 x
Rotor Radius (cm) x (RPM)
Drucker Diagnostics | +1-814-692-7661 | CustomerService@DruckerDiagnostics.com | 12
G-F
ORCE
2
Radius
6pl Rotor: 6 tubes up to 35 mL
4pl Rotor: 4 tubes up to 35mL
15 in x 17 in x 9 in
(38 cm x 43 cm x 23 cm)
39 lbs. (17 kg)
59 dB A
16 – 32 °C
100 -240 VAC
50/60 Hz
415W (MAX)
½ H.P. Brushless DC
1-30 (+/- 2%)
2,500 RPM (+/- 100)
140.0
grams
In Inches:
RCF or G-force = 0.0000284 x
Rotor Radius (in) x (RPM)
6 in. (15.3 cm)