Malfunction and troubleshooting
Malfunction
Device is not operational
Pedal is not functioning
Suction pump is not
functioning
Suction pump is not work-
ing properly
If a fault cannot be rectified, please contact your supplier or an authorized service center. The addresses are provided on
the last page of tis operating instructions.
16
Cause
Pump is not switched on
No connection to electricity-
supply
Wrong voltage
Defective fuses
On/Off-AIR-Pedal is not
connected
Control unit is not
switched on
Incorrect operation
On/Off-AIR-Pedal is not
connected
Vacuum-system is not air
tight
Air inlet port at rear of the
device is open
Tubes are connected wrong
Jar is full and Overflow pro-
tection has locked down
Incorrect operation
Vacuum controller is not
opend wide enough
Vacuum-systemis not air tight Check all seals and tubes.
Solution
Switch main switch "I/O"
to "I"
Connect power cord to
electricity-supply
Check power supply of your
Pump
Replace fuse
Connect On/Off-AIR-Pedal
with device at rear
Switch main switch "I/O"
to "I"
Read instruction manual
carefully
Connect On/Off-AIR-Pedal
with device at rear
Check all seals and tubes.
Make sure char lid is properly
closed
Close air inlet port with its
cap.
Connect tubes correctly
Replace full jar by a fresh,
empty jar
Read instruction manual
carefully
Turn Vacuum controller
clockwise
Make sure char lid is properly
closed
Operating instructions Vacuson 40/60, Ref. 31997, V43/13
Vacuson 40/60
Reference in manual
7.1 Switching device on and off
6.4 Device preparation
5.0 Device overview
9.1 Replacing the control unit fuse
6.4 Device preparation
7.1 Switching device on and off
6.4 Device preparation
6.3 Preparation of secretion jar
6.4 Device preparation
9.7 Seals and tubes
6.4 Device preparation
6.4 Device preparation
7.5 Emptying secretion jar
7.4 Regulating suction process
6.3 Preparation of secretion jar
6.4 Device preparation
9.7 Seals and tubes