MEGABRAS MPK-253 Guia Del Usuario página 8

Publicidad

Idiomas disponibles

Idiomas disponibles

WAIT...
This message appears each time the equipment needs to adjust any parameter
in order to optimize the readings.
PRESS START
The equipment is ready to start a measurement, thus the operator has to press
the START
button.
LOW CURRENT
It shows that the test current is not enough to carry out the reading. It appears at
the beginning of each test and it keeps on being there up to the operator rotates
the TEST CURRENT
control clockwise, as necessary for the test current to
be higher than 10 % of the nominal current in the scale. The inappropriate con-
nection of the cables may cause a difficult circulation of test current. If this mes-
sage keeps on being displayed, please check that the current cables are con-
nected appropriately.
OVERRANGE
It indicates that the measured resistance is higher than the maximum value
readable in the selected scale.
H
Indicates that the value is the one retained in the memory when pressing the
HOLD
button.
BAT
It shows that the battery is quite discharged. The available energy is enough to
work during a few minutes only. It is necessary the battery to be recharged.