4. Design and function
4.2
Integrated data logger
The gas density indicator features an integrated data logger.
This data logger can be turned on or off via [MENU] / [Logger] or via the
"myWIKAdevice" app under [Logger] / [Upload].
EN
4.3
Voltage supply
Three AA alkaline batteries are used as the voltage supply for the instrument. These are
included in delivery.
The battery life time is up to 2,000 hours for continuous operation (without backlighting
and with WIKA-Wireless deactivated).
In the upper right half of the display there is a symbol for the battery capacity. Instructions
on batteries (see chapter 8.2 "Battery replacement").
The battery indicator lights up
To avoid false readings, replace the batteries.
Instructions on batteries (see chapter 8.2 "Battery replacement").
4.4
WIKA-Wireless
To start wireless data transfer, you must set it to "On" under "Menu / Basic settings /
Wireless". Once this has been done, the wireless symbol blinks in the display. As soon
as the gas density indicator is connected to a smartphone via this wireless interface, the
symbol lights continuously.
Under "Menu / Basic settings / Wireless / Wireless", a distinction can be made
between the communication type WIKA-Wireless Classic or WIKA-Wireless Classic with
WIKA-Wireless Low Energy (= LE).
WIKA-Wireless Low Energy (= LE) is required to communicate with an IOS-enabled
mobile phone via the myWIKA-device smartphone app.
It is recommended to select WIKA-Wireless Classic for a connection with PC and/or an
Android-enabled instrument.
4.5
Protective rubber cap for case
Optionally, the gas density indicator can be fitted with an impact-resistant protective
rubber cap.
16
WIKA operating instructions, model GDI-100-D