Descargar Imprimir esta página

WIKA CTD9100-165 Manual De Instrucciones página 23

Calibradores de temperatura/microbanos de calibración

Publicidad

Idiomas disponibles

Idiomas disponibles

Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
Fig. 43: Cooling gradient entry
Press the P key.
SLoF flashes on the upper PV display.
The respective currently set cooling gradient
appears on the lower SV display.
Fig. 44: Display of the cooling gradient input
Press the 5 key to increase the cooling gradient
SLoF.
Press the  key to decrease the cooling
gradient SLoF.
Press the 5 and  key to raise and
lower the value by 0.1 respectively.
If the keys are held pressed for at
least one second, the value increases
or decreases quickly and after two
seconds even more quickly; this
means the desired value can be
reached rapidly.
Press the P key to confirm the set
cooling gradient SLoF.
The display returns to the parameter
level and other parameters can be
set.
If no key is pressed for approx. 15
seconds, the device automatically
returns to a previous level up to the
calibration mode.
After carrying out the settings, the
calibrator/micro calibration bath uses
the new values only when modifying
the set temperature or switching the
calibrator/micro calibration bath off
and on again.
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
6.3.7.3 Setting the duration time
The duration time "dur.t" is active if the set
temperature SP1 has been achieved. Subsequently,
the calibrator/micro calibration bath automatically
switches to set temperature SP2.
The setting range extends from 99:59 [hh:min] to
00:00 [hh:min].
The function is deactivated if dur.t
= InF (In no Function) has been set.
You are in the parameter level (as described in
section 5.3.7 ).
'rEG appears on the upper PV display.
SLor flashes on the lower SV display.
Fig. 45: Parameters for heating gradient SLor
Use the 5 or  key to select the parameter dur.t.
'rEG appears on the upper PV display.
dur.t flashes on the lower SV display.
Fig. 46: Parameters for the duration time dur.t
Press the P key.
dur.t flashes on the upper PV display.
The respective currently set duration time
appears on the lower SV display.
Fig. 47: Duration time entry
GB
23

Publicidad

loading

Este manual también es adecuado para:

Ctd9100-450Ctd9100-650Ctb9100-165Ctb9100-225