Service Information
7
This service information describes the connection of the IntelliBridge to the Maquet CARDIOSAVE
pump.
Data communication
The IntelliBridge EC10 module supports LAN data communication with the following
communication settings on the Maquet CARDIOSAVE pump.
• IP / subnet:
• Default port:
• Protocol:
• Socket type:
The network configuration should only be accessed by authorized personnel.
Instruction for connection
Connect the IntelliBridge EC5 RJ45 LAN connector to the LAN connector of the Maquet
CARDIOSAVE pump as shown below .
IntelliBridge EC5 connection to LAN port
1
2
3
class C network address like: 192.168.100.50 with subnet 255.255.255.0
33320
TCP/IP – HTTP 1.1
Server
1
Maquet CARDIOSAVE pump
IntelliBridge EC5 ID module
EC5 device label: "Maquet CARDIOSAVE"
• HIS/CIS server:
• Network connection: HIS/CIS
2
7
192.168.100.2
3
25