Inflating Further; Discontinuing A Measurement; Memory - Displaying The Last Measurement - Procheck UAM0XB-H Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

e) When the measurement has been concluded, a long
peep tone sounds. The measured systolic and diastolic
blood-pressure values as well as the pulse frequency
are now displayed.
The cuff pressure can now be quickly released by pres-
sing the release valve at the pump ball.

5.5. Inflating further

If the cuff is not inflated sufficiently, the measurement is
broken off after a few seconds and a flashing arrow
pointing upwards appears in the display.
It is then necessary to inflate further to at least 20 mmHg
higher than the previous value.
The instruction to inflate further can appear several times
if the pressure is still not sufficient.
Attention:
If the cuff is inflated too far (over 300 mmHg), «HI» appears
in the display and a warning signal is sounded. The cuff
pressure must then be immediately released using the quick-
release valve) at the pump ball)!
As soon as the cuff is inflated sufficiently and the measure-
ment is proceeding, continue as described under points d–e.

5.6. Discontinuing a measurement

If it is necessary to interrupt a blood pressure
measurement for any reason (e.g. the patient
feels unwell), press the quick release valve at
the pump ball. The device then immediately lo-
wers the cuff-pressure automatically.
5.7. Memory – Displaying the last measurement
The blood-pressure monitor automatically stores
the last measurement value. By pressing the
MEMORY button for at least 3 seconds, the last
measurement will be displayed.
Further information
Measurements should not occur soon after each other, since otherwise, the results will be falsified. Wait
therefore for at least 1 minute in a relaxed position, sitting or lying, before you repeat a measurement.
MR
8

Publicidad

Tabla de contenido
loading

Productos relacionados para Procheck Procheck UAM0XB-H

Tabla de contenido