Specifications
Model:
Operation principle:
Display:
Measuring range:
Blood pressure
measurement:
Laboratory accuracy:
Clinical accuracy:
Inflation:
Deflation:
Power supply:
Operating temperature:
Storage temperature:
Humidity:
Cuff:
Clinical validation:
(Please note that a validation against intra-arterial measurements can lead to different
results for devices which are calibrated against auscultatory reference.)
Type BF equipment
Subject to change without notice.
This appliance conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
DIN EN 1060-1: 12/95, AMD 1 09/02 «Non-invasive sphygmomanometers» –
Part 1: General requirements
DIN EN 1060-3: 9/97 «Non-invasive sphygmomanometers» –
Part 3: Supplementary requirements for electro-mechanical blood
pressure measuring systems
This product conforms to the provisions of the EC directive 93 /42 / EEC
(Medical Device Directive).
0297
12
BP 2510
Oscillometric method
Liquid Crystal Display
0 – 300 mmHg (cuff pressure)
40 – 160 / min (pulse rate)
50 mmHg (minimum diastolic value)
250 mmHg (maximum systolic value)
± 3 mmHg (cuff pressure)
± 5 % of reading (pulse rate)
According to AAMI-SP 10 with auscultatory reference:
< 5 mmHg systematic offset
< 8 mmHg standard deviation
Automatic, adapted to blood pressure
Electronic control valve adapted to heart frequency
2 batteries, type LR 03 (AAA)
(e.g. Duracell Ultra)
+ 10 °C to + 40 °C
– 20 °C to + 60 °C
Up to 85 % RH
Suitable for wrist circumference of 13.5 to 20 cm
According to AAMI-SP10 with auscultatory reference.