INSTRUCTION MANUAL CODE 80301 REV 0
I D IO M A
FECHA/HORA
INTRODUCIR
B R I L L O
CONTRASEÑA
HOME
Fig 1. Programming screen
Selected program.
PROGRAM 00
TEMPERATURE: 16.0ºC
TIME: 00:00
UNDERTEMP: 0ºC
OVERTEMP: 0ºC
HOME
SETUP
Fig 7.3. Programming screen.
J
.P. SELECTA s.a.
7.2 Prgramming screen.
UNIDADES
7.2.1 Set time and date:
C AMB I AR
Press on Date/time button.
CONTRASEÑA
Press over push buttons to set time and date.
Use Up and down icons to entry a value.
C AL
7.2.2 Set temperature, language and screen contrast.
In a similar way set the temperature, languaje and screen contrast.
7.2.2 working parameters protected under password.
Some parameters setting are password protected.
Temperature calibration.
.. ..
7.3
Up to 8 programms of temperature and tiem can be stored.
Each program have the following parameters:
Te mpe ra tur e:
Time:
Un de rt emp:
O ve rt e mp :
Programm
From main menu press over «PROGRAMMING»
selector.
Fig7.3 screen is displayed.
Select a program (from 00 to 07) with keys:
SEL
Set working temperature and the other parameters:
Press:
SAVE
keys
HH:MM:SS DD/MM/YY
In a similar way press:
TIME: 00:00
UNDERTMP: 0ºC
OVERTMP: 0ºC
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Feb/2015
(It can be modified without notice)
Temperature and working time setting .
Temperature: from ambient+5ºC ... 200ºC
HH:MM
Alarm value for temperature drop.
Alarm value for overtemperature.
TEMP: 16.0ºC
will be shown to select temperature.
Set de working time.
A time setting value of 00.00 means «continuous
operation».
Sets the undertemperature alarm value.
Sets the overtemperature alarm value.
Press
SAVE
to save .
Tel 34 937 700 877 Fax 34 937 702 362