Cuvette usage and handling
It is important to maintain a standard for cuvette usage
as their quality will directly impact the results of your
experiments. If the cuvette has fingerprints, smudges, or
cracks the sensor will detect them and give off an
improper colorimeter reading. Smudges left from
handling can easily be wiped off with tissue paper.
Proper usage:
1. Carefully remove the cuvettes from their packaging.
2. Load your liquid sample into the cuvette before
placing it into the sensor to prevent spilling and
damaging the sensor.
3. Plug the cap into the cuvette and gently invert a few
times. Shaking will create air bubbles which will be
detected by the sensor and give an incorrect
reading.
4. Before placing the cuvette in the sensor; hold it as far
towards the top as you can and wipe off the bottom
with a tissue. Do not touch the bottom of the cuvette
again as this will leave a residue which could affect
the results of the experiment.
V2018.01
19