Descargar Imprimir esta página

Balluff BTL7-A-K Serie Manual De Instrucciones página 43

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 109
BTL7-A/C/E/G5 _ _ -M _ _ _ _ -HB/WB-KA_ _/K_ _/FA_ _/F_ _
Micropulse Transducer - Rod Style
6
Calibration procedure (continued)
6.3.2
Adjusting
The detailed procedure for adjusting is
described on page 17 ff.
The factory-set null point and end point are replaced by a
new start point and end point and the associated output
values can be adjusted. The start and end values can be
adjusted as you like up to the limits.
Steps
► Move magnet to the new start position.
► Read new start point by activating the programming
inputs.
► Set the new start value by activating the programming
inputs.
New start value
New null point
Fig. 6-4:
Adjust a new start value
www.balluff.com
► Move magnet to the new end position.
► Read new end point by activating the programming
inputs.
► Set the new end value by activating the programming
inputs.
New end value
Before
Fig. 6-5:
Adjust a new end value
After
6.3.3
Reset
Restoring the transducer to its factory settings.
The detailed procedure for the reset is
described on page 19.
Before
After
New end point
english
15

Publicidad

loading