2 - Check that the protective silicon sleeve correctly covers all the metallic surface of the contra-angle.
3 - Open the rubber cover and plug the data transfer cable to Dual Move.
4 - Power ON Dual Pex and connect the data transfer cable to Dual Pex. CONNECTED ! is displayed on the Dual Move's
screen when the connection is correctly done.
11. Connection testing (Fig. 11)
Fig. 11.A Insert a fi le into the contra-angle, create a short circuit between the lip clip and the fi le.
The fi le may start rotating if "Auto start / stop" function (section 13, Fig. 13.C). is activated.
Fig. 11.B Press the Dual Move power switch. A full-length white bar appears next to the line representing the canal, indicating that the
system is working properly and is ready to be used.
Hang the lip hook on the patient's lower lip and start the treatment.
12. Dual Move screen during treatment (Fig. 12)
Fig. 12
The white strip on the Dual Move screen shows the progression of the fi le into the canal.
The beeps frequency increases when the fi le approaches the apex.
13. Combination functions (Fig. 13)
The following advanced functions are only usable when Dual Move is connected to Dual Pex.
1 - Touch < / > to select a memory number (M1...M6)
2 - Touch the S key for more than one second to enter the advanced function setting interface.
3 - Touch < / > to activate (ON) or deactivate (OFF) the advanced functions. If "CONNECT FUNCTION" is deactivated
(OFF), all the advanced functions will be inactive, even if they are individually activated (respective status set to "ON").
The code of following function to be set (APA, ASS, ASD) is displayed at the bottom of the screen, just next to the
battery status.
4 - Touch S to switch to the following combination function.
5 - Touch < / > to activate (ON) or deactivate (OFF) the function then S to go to the next function.
Automatic return to the standby screen if no action for a while (time-out to be set, factory value 5s).
Connection functions (CONNECT Function)
When Dual Move and Dual Pex are connected, set "ON"
13.A
to have the possibility to use the combination functions,
depending on their status.
Apical auto-reverse (APA)
Automatic reverse when the fi le reaches the reference
13.B
point (*). Automatic restart as soon as it comes back
above the reference point.
Factory status of advanced function for each Memory:
Function
Apical Reverse (APR)
Auto Start & Stop (ASS)
Apical Slow Down (ASD)
(*) Position of the Reference point (RP) is set through the Dual Pex interface then automatically sent to Dual Move and
displayed as a cursor on the Dual Move screen. When the fi le reaches the reference point, Dual Move will start the
APICAL Reverse, APICAL Slow Down depending on their respective status (ON / OFF).
13.C
13.D
M1, M2, M3
Activated (ON)
Activated (ON)
Deactivated (OFF)
Auto Start stop (Auto Stt Stp - ASS)
When the fi le enters the orifi ce of the canal, the motor will
automatically start. When the fi le leaves the orifi ce of the
canal, it will automatically stop.
APICAL Slow Down (ASD)
Automatic deceleration once the fi le exceeds the reference
point (*).
M4, M5, M6
Deactivated (OFF)
Activated (ON)
Activated (ON)
7