Calibration
Calibration verification
Midpoint and endpoint system method:
1. Place the laser in the middle of a room with 5 metres between 2 of its
walls.
2. Turn on the laser facing wall 1(P1) and mark the horizontal line (L1).
3. Rotate the laser 180° from its initial position without moving it from the
spot. The laser will face wall 2 (P2), the opposite wall to wall 1. Mark the
horizontal line (L2) on this wall 2.
4. Place the laser next to wall 1 and align the horizontal beam at the exact
height of the first marked line (L1).
5. Turn the laser 180°, projecting the horizontal light on wall 2. Mark the
horizontal line (L3) again on wall 2.
6. If the distance between L2 and L3 (d) is more than 2mm you must
calibrate your laser.
7. You can repeat this same system with the vertical axes.
Step 1 and 2