Calibration Menu
You can choose the automatic or manual calibration mode, in both cases the calibration at pH 7 is automatic.
-
Automatic calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press
will show the countdown of the 60 seconds required to complete the calibration. If the quality of the alignment is
less than 50%, an error appears on the display; press
automatically exit the calibration); if the quality is higher than 50%, the value appears on the display; pressing
you will be prompted for the pH 4 or 9 buffer solution; at this point the procedure is similar to the previous
calibration.
-
Manual calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press
will show the countdown of the 60 seconds required to complete the calibration. If the quality of the alignment is
less than 50%, an error appears on the display; press
automatically exit the calibration); if the quality is higher than 50%, the value appears on the display; pressing
on the display will flash the pH 7.00 value; press
and start the calibration procedure as described previously.
EM00136304
to insert the value of my solution, then press
DosingPump.ir
rev. 1.0 EN
to exit the calibration (after 4 seconds the pump will
to exit the calibration (after 4 seconds the pump will
and the display
and the display
to confirm