Use at lower temperatures
Due to heat input from the pump and stirrer motor, the minimum operating temperature without
accessory cooling depends on the size of the bath. The smallest bath, P5, S5, have an
operating temperature of approximately 10C above ambient without a lid (15C with a lid).
The other baths can be used at temperatures of 5C above ambient without accessory
cooling.
For lower temperature operation, accessory cooling is required.
3.
Operation
3.1.
Controls and indicator lamps
3.1.1. Front Panel
Heater lamp
Alarm lamp
Display
Function, F
Navigation
control
Power
3.1.2. Power
The power (mains) switch is located in the lower left hand corner of the front
17091 Version 12 - October 2009 75T
Page 7
www.grant.co.uk
Set overtemperature °C
Push to reset
Set, S
panel.
GR150, GP200
Operating Manual