Cable management
Do not bend, stretch, or tightly wrap the LoFlo Module cable around objects such as a bedrail, your hand, or the sensor itself. Do not fold
the cable over on itself and secure with a rubber band. The sensor cable should be stored ONLY in a minimum 3 inch (7.6 cm) diameter
coil and secured with a cable tie as shown.
NOTE: Any serious incident that has occurred in relation to this device should be reported to Respironics Novametrix and the competent
authority of the country in which the user and/or patient is established.
Disposal
Dispose of in accordance with all applicable Federal, State, and local Medical/Hazardous waste practices.
Specifications
CO
accuracy
2
CO
measurement range
2
CO
noise
2
CO
resolution
2
CO
stability
2
Data Sampling Rate
Gas Sample Rate
Respiration rate accuracy
Respiration rate range
0 - 40 mmHg
±2 mmHg
41 - 70 mmHg
±5% of reading
71 - 100 mmHg
±8% of reading
101 - 150 mmHg
±10% of reading
Above 80 BPM
±12% of reading
0 to 150 mmHg, 0 to 19.7%, 0 to 20 kPa
RMS noise of the sensor is ≤ to 0.25 mmHg at 5% CO
0.1 mmHg
0 to 69 mmHg
0.25 mmHg
70 to 150 mmHg
Short term drift over four hours shall not exceed 0.8 mmHg maximum.
Long term drift accuracy will be maintained over a 120 hour period.
100 Hz
50 ml/minute ±10 ml/minute
±1 breath
2 to 150 breaths per minute (BPM)
2
> 3 inches (7.6 cm)
3