6. Battery condition
The charging condition of the battery can be verified before or during the
resistance measurement. In order to achieve that, you have to press the
Battery charge condition
well charged the indicator pointer will stop over the blue zone of the arc. If
the pointer stops over the red zone, it means that the battery is
discharged and so it has to be quickly charged. If during measurement
the charge of the battery achieves a critical level, the display will show the
BAT message notifying that the charge level is low. After a few minutes
the measurement will be automatically interrupted in order to preserve the
battery from a deep discharge that is prejudicial for its expected useful
life.
6.1. Battery and charger
Battery description
The
MEGABRAS
rechargeable, lead-acid, 12V - 7Ah battery. At the end of the useful life,
that battery has to be recycled or disposed of properly in order to take
care of the environment.
Battery charger
The built-in battery charger is always active when the equipment is
connected to the mains supply, even if the On/Off
Charging procedure:
Check that the On/Off
Connect the equipment to the mains supply.
The indicator
the charge. At that point, it will change to a green light, being like this up
to the equipment disconnection from the mains supply.
key while the equipment is turned on. If it is
MPK-204
micro-ohmmeter
key is OFF.
will keep on lightning with a red light up to completing