TROUBLESHOOTING

Use the following table to help identify possible causes and corrective actions for problems you may experience. If you need further assistance, contact your M Service
Center or call M Technical Service in U.S.A., 1-800--60. In Canada, call Technical Service at 1-800-67-1.
Problem
Unit doesn't respond when switched on
Battery doesn't last long enough
Very high reading(s).
WIBGET monitor sends data but PC does not receive.
Computer produces nonsense.
Computer locks up
If an 'E'Code appears on the display (even momentarily) during POWER ON/SELF DIAGNOSTICS, refer to the following:
Code Displayed
E1, E, E
E, E5
E6GT, E6WB, E6DB
E7, E8

PRODUCTS, ACCESSORIES AND PARTS

Never alter or modify this instrument. Substitution of components may impair intrinsic safety. Doing so may adversely affect product performance and result in sickness or death.
Repair or replace parts only with the M components approved for this unit. Failure to do so may adversely affect product performance and result in sickness or death.
WIBGET Heat Stress Monitors
Part #
Description
RSS-1
M
RSS-1DL
M
Replacement Parts and Accessories
Part #
Description
-10
M
-10
M
-500
M
-100
M
-505
M
7-0500
M
9-010
M
-0
M
-00
M
-01
M
-0
M
-0
M
-0wp
M
-0wp
M
Possible Cause
Battery/charger failing.
Battery/charger failing.
Connector wet.
RSS-1 internal switch # setting incorrect.
RSS-1 internal switch #1 setting incorrect.
RSS-1 internal switch # setting incorrect.
Possible Cause
Component failure
Analog O/P out of calibration
Sensor specific malfunction
Memory failure or low battery
WIBGET
Heat Stress Monitor
WIBGET
Heat Stress Monitor with Data Logging
Dry Bulb Sensor
Wet Bulb Sensor
Wet Bulb Kit (contains demineralizer, 6 wicks &  sponges)
Globe Sensor
9 Volt NiMH Battery and Charger
9 Volt NiMH Battery
Water bottle
Personal Sensor Set (with hardhat clip attachment)
10 ft Sensor Assembly Cable
0 ft Sensor Assembly Cable
50 ft Sensor Assembly Cable
100 ft Remote Sensor Assembly Cable
50 ft Weatherproof Remote Sensor Assembly Cable
100 ft Weatherproof Remote Sensor Assembly Cable
Possible Solution
Replace battery
Replace battery
Rinse connector with iso-propyl alcohol and dry.
Reset switch # then reboot PC to reset com port.
Set to alternate position (switch baud rates are 00 or 100).
Set to alternate position.
WARNING
1
Possible Solution
Return for repair
Adjust Trim R to 50 mv at 5.0°C.
Plug in, clean or replace affected sensor(s)
Return for repair, or recharge/replace battery