Accuracy Table
Transferpette
Volume range
µl
10
20
25
50
100
200
500
1000
2000
* A = Accuracy, CV = Coefficient of Variation
20 °C
!
Ex
Final test values related to the nominal capacity (maximum volume)
indicated on the instrument obtained when instrument and distilled
water are equilibrated at ambient temperature (20 °C/68 °F) and with
smooth operation. According to DIN EN ISO 8655.
Note:
The device is marked in accordance with the German Measurement and
Calibration Act as well as the Measurement and Calibration Regulation:
18
Character string: DE-M (DE for Germany), framed by a rectangle, as well
as the last two digits of the year in which the marking was affixed (here:
2018).
38
S , Fixed volume
®
A*
CV*
≤ ± %
≤ %
1
0.5
0.8
0.4
0.8
0.4
0.8
0.4
0.6
0.2
0.6
0.2
0.6
0.2
0.6
0.2
0.8
0.3
Recommended
type of tip, µl
0.5 -
20
2 -
200
2 -
200
2 -
200
2 -
200
2 -
200
50 - 1000
50 - 1000
500 - 5000