I I I I I Temperature influence on measurement accuracy (for all functions)
±0.15 % per °C
I I I I I Altitude: < 2000 m
7.5
Mechanical specifications
Outside dimensions of casing (L x W x H): 240 x 185 x 110 mm
I
Weight : approx. 3.4 kg
I
7.6
Conformity to international standards
Electrical safety according to: EN 61010-1, EN 61557
I
Double insulation:
I
Pollution level: 2
I
Installation category: III
I
Max. voltage in relation to earth: 600 V
I
7.6.1. Electromagnetic Compatibility
Emissions and immunity in an industrial setting compliant with EN 61326-1.
I
7.6.2. Mechanical protection
IP 54 according to NF EN 60529
IK 04 according to NF EN 50102
For maintenance, use only specified spare parts. The manufacturer will not be held responsible
!
for any accident occuring following a repair done other than by its After Sales Service or approved
repairers.
8.1.
Upkeep
8.1.1. Replacing the batteries (C.A 6541)
The battery voltage comes up on the small digital display for 2 seconds when switching on the instrument.
The main display indicates "bAt". This test is made internally on a load corresponding to that during
functioning.
Before making the measurement, make sure that the
the switching on stage the battery voltage appears for 2 seconds on the small display).
If this is not the case, it is essential that you should change all the batteries, taking all the necessary
precautions whilst opening the instrument.
Check that none of the terminals is connected and that the selector switch is definitely on OFF
!
before accessing the battery compartment found inside the instrument.
Exact type of batteries specified: LR14 (alkaline)
To open up the instrument, unscrew the 4 captive screws situated on the underside of the case. As soon
as they turn freely, place the instrument on a table and press on the screws to extract the front panel.
Turn over the instrument to entirely remove the front panel unit from the casing. The batteries are then
accessible at the back of the panel unit after having unscrewed the battery compartment's two screws.
Take care not to damage the seal during removal and replacement of the front panel unit.
8. MAINTENANCE
symbol is not flashing on the screen. (after
59