11. SERIAL OUTPUT
The serial output of the balance uses the EIA RS232 standard
Balance and computer/ peripheral settings must be the same.
11.1. SERIAL OUTPUT CONNECTION
The connection of the balance is:
Balance
RS 232 out
RS 232 in
GND
11.2. DATA TRANSMISSION FORMAT
The format of the data transmitted by the balance is different as a function of the selected protocol by the
menu option 9.2.1.3.
EURO and CRYSTAL protocols are suitable for computer interfacing, the PRINTER protocol may be used to
drive external printers.
The answer to any type of command is formatted with a measure field, a unit field and a status flag field as
follows:
D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 B
1
2
3
4
D0 ...D9
Measure
B
Blank
U
Unit
F1, F2
Status
CR
Carriage Return
LF
Line Feed
The status flags values of F1 are:
D
Transmitted data is valid
O
Balance in over-range
U
Balance in under-range
T
Tare in progress
C
Calibration in progress
Z
Zero acquire
I
Init system
The status flags values of F2 are:
S
Measure is stable
I
Measure is not stable
E
Measure is in error
A
Automatic calibration ackowleged
B
Calibration in progress
Page 30 of 45
DB9
CRYSTAL Protocol
5
6
7
8
D25
2
3
5
U
U
9
10 11 12 13 14 15 16 17 18 19
with sign & decimal point, right justified
Space between measure and unit
Measuring unit
Flags indicating the status of the balance
ASCII code <CR>
ASCII code <LF>
INSRUCTION MANUAL - 60.50.630 9th issue –December 2014
CRYSTAL analytical balance
3
2
7
U
B
F1 F2 CR LF