Satelec Suprasson P5 Booster Manual De Uso página 10

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 25
F
AULT
No operation
No spray
No water at tip but
vibration present
Insufficient power
Low vibration
No ultrasonic output
Water leak between
handpiece and cord or at
front/back junction of
handpiece
P
OSSIBLE CAUSE
• Defective mains cord connection
• Presence of fluid or moisture between
handpiece and cord
• Blown fuse
• Defective water tube connection
• No water pressure
• Clogged filter or defective solenoid
valve
• Obstructed tip or file
• Wrong tip selection
• Incorrect spray adjustment
• Worn or deformed tip
• Incorrect use : wrong angle of
application or inadequate pressure on
tooth
• Presence of fluid or moisture between
handpiece and cord
• Incorrect tip tightening
• Defective connector contact
• Cut wire in handpiece cord
• Worn sealing gasket (1,15 x 1) in
handpiece
10
R
EMEDY
• Check mains socket
• Dry any moisture,
especially on electrical
contacts
• Return to SATELEC After
Sales Service
• Check water supply
• Check pressure
• Clean or replace filter
• Return to SATELEC After
Sales Service
• Remove obstruction
• Check tip
• Adjust spray
• Replace tip
• See tip utilization method
• Thoroughly dry electrical
contacts
• Tighten with wrench
• Clean cord contacts
• Return to SATELEC After
Sales Service for cord
replacement
• Replace seal

Publicidad

Tabla de contenido
loading

Tabla de contenido