Recommended separation distances between portable and mobile RF communi-
cations equipment and the Interactive Building Block Compressor Nebulizer.
The Interactive Building Block Compressor Nebulizer is intended for use in an electromag-
netic environment in which radiated RF disturbances are controlled. The customer or the
user of the Interactive Building Block Compressor Nebulizer can help prevent electromag-
netic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Interactive Building Block Compressor
Nebulizer as recommended below, according to the maximum output power of the com-
munications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance din metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where Pis the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propa-
gation is affected by absorption and reflection from structures, objects and people.
Rev.2.08.10.15
14
............................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL MQ7000, MQ7000EU
............................................................................................................
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √ P
0.12
0.38
1.20
3.79
12.00
m
80 MHz to 800 MHz
d = 1.2 √ P
0.12
0.38
1.20
3.79
12.00
800 MHz to 2,5 GHz
d = 2.3 √ P
0.23
0.73
2.30
7.27
23.00