Sensor de control de temperatura

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 12
weaken the sensor adhesive, damage the sensor, or compromise the device
performance.
• Avoid placing the sensor in the center of the forehead as this may affect
sensor accuracy.
7.
To reduce the risks associated with exposure to biohazards:
• Follow facilities policies and procedures for disposal of contaminated
materials.
• Always perform the decontamination procedure prior to returning the
SpotOn temperature monitoring system for service and prior to disposal.
8.
To reduce the risks associated with entanglement:
• Do not leave pediatric patients unattended while using the SpotOn
temperature monitoring system.
CAUTIONS
1.
To reduce the risks associated with skin maceration:
• Do not use the sensor on damaged or compromised skin.
2.
To reduce the risks associated with cross-contamination:
• Clean the sensor cable before connecting to a new sensor.
3.
To reduce the risks associated with impact and facility medical device damage:
• Do not use the SpotOn control unit as a handle or transport or to move the
device to which it is attached.
• Do not initiate temperature monitoring unless the SpotOn control unit is
safely placed on a hard, flat surface or is securely mounted.
4.
To reduce the risks associated with environmental contamination:
• Follow applicable regulations when disposing of this device or any of its
electronic components.
NOTICES
1.
The SpotOn temperature monitoring system meets medical electronic
interference requirements. If radio frequency interference with other equipment
should occur, connect the unit to a different power source.
2.
Federal law (USA) restricts this device to sale by or on the order of a licensed
healthcare professional.
3.
To avoid SpotOn temperature monitoring system damage that may impact
performance:
• Do not store the SpotOn control unit and system components in a wet or
damp place.
• Do not spray cleaning solutions onto the control unit or into the sensor
cable connector.
• Do not immerse the SpotOn temperature monitoring control unit or system
components in any liquid or subject them to any sterilization process.
• Do not use solvents such as acetone or thinner to clean the control unit;
avoid abrasive cleaners.
• Clean control unit exterior with a damp soft cloth and 70% mixture of
isopropyl alcohol and water.
• Do not immerse any of the devices or use a dripping wet cloth for cleaning.
4.
The SpotOn temperature monitoring sensor is not made with natural latex
material.
5.
To the full extent permitted by law, the manufacturer and/or importer declines
all responsibility for injury resulting from the unit being used in conjunction
with unapproved system components.