Technical specifications
Test currents
Resistance ranges
Resolution
Output voltage
Measurement principle
Basic accuracy
Advanced features
Serial data output
Environmental protection
Safety class
E.M.C
Electrostatic immunity
Power supply
Built-in battery charger
:
1 mA - 10 mA - 100 mA - 1 A - 10 A. Each current may
be continuously adjustable from 0 to 100%.
:
0-2000 µΩ @ 10 A
0-20 mΩ @ 10 A
0-200 mΩ @ 1 A
0-2000 mΩ @ 100 mA
0-20 Ω @ 10 mA
0-200 Ω @ 1 mA
:
0,1 µΩ @ 10 A
:
Up to 10 Vc.c. @ 1 A (open circuit)
:
Four-terminal, Kelvin-type.
:
±(0,10 % of reading + 0,005 % full scale)
:
Digital direct reading of very low resistances in the
alphanumerical display, with up to 4½ digits. Very fast
and accurate measurements.
:
RS-232 @ 4800 bps. Suitable for data collection in an
external serial printer, computer or data-logger.
:
IP54 with closed lid.
:
Meets the requirements of IEC61010-1/1990, IEC
61010-1/1992 amendment 2
:
In accordance with
:
In accordance with IEC 1000-4-2
:
Internal battery powered. Rechargeable battery, sealed
lead, 12V - 7Ah or mains.
MPK-2000e is manufactured in different versions, to be
:
compliant with the mains supply specifications in the
destination country. Please, check the value indicated
in the Control Panel in order to be sure that it is appro-
priate for the local A.C. supply.
14
IEC 61326-1