1.
Place the rolled Gelli-Roll
time needed for the pad to reach the desired temperature.
Connect the male coupling and the female coupling from the Cat#286 hose to the unit. Make sure they "click" as they lock
2.
together.
Attach the pad to the Cat#286 hose by inserting the pad connectors into the hose connectors. Connectors should "click" as they
3.
lock together.
4.
Check the equipment reservoir to ensure it is full of sterile water or water that has been passed through a filter less than or
equal to 0.22 microns. Turn the unit on and select the desired temperature setting. The unit should be turned on approximately
45 minutes prior to first use, preferably first thing in the morning.
Setting of temperature is dependent upon patient's core temperature and patient's clinical condition. This temperature should
5.
be adjusted as required to attain and/or maintain target core temperature according to the physician's order.
6.
After the allotted time, unroll the Gelli-Roll
Check pad before each use for product integrity (i.e. no leaks, flow is normal, etc.).
7.
If cooling mode is selected, cool-down will take approximately 40 minutes.
8.
After filling and positioning, make sure the pad and hose are free of kinks that may restrict flow.
•
Follow the thermal regulating system operation manual for operating instructions.
•
To discontinue therapy: turn the regulating system off. The water in the pad will automatically drain back into the unit. Allow it to
fully drain before disconnecting the pad from the unit or removing from bed.
•
For cleaning and disinfecting, always use conventional hospital-approved topical equipment cleaners and disinfectants that do not
contain alcohol. Pads are washable at temperatures of approximately 108°F (42.2°C). Make sure pads do not enter washing
machines or dryers. Avoid alcohol and other strong, undiluted disinfectants. These may cause staining or hardening of the pad's
outer skin. Thoroughly rinse product with clear water to remove any residue from cleaning solutions. A solution composed of 1:10
parts bleach to water can also be used for cleaning purposes.
Note: Do not use gas sterilization or autoclaving for cleaning and disinfecting this product.
®
on the bed, stretcher or OR table. Unrolling the pad prior to the warming period may increase the
®
. Ensure that the pad lays flat with the hose routed without kinks towards the unit.
OPERATING INSTRUCTIONS
SET UP
CLEANING