The above listed models are intended for use in the electromagnetic environment specified below. The customer or the user of
the above listed models should assure that they are used in such an environment.
Immunity test
Electrical fast
transient/bursts
IEC 610004-4
Surges
IEC 61000-4-5
± 0,5kV, ±1 kV , ± 2 kV line(s) to ground
Conducted RF induced
by RF fields
IEC 61000-4-6
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Comment: n) The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to
13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz
to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to
18,17 MHz,21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
Input a.c. power PORT
Test Condition
± 2 kV
100kHz Repetition frequency
± 0,5kV, ±1 kV line(s) to line(s)
3 V
0,15 MHz – 80 MHz
6 V m) in ISM and amateur
radio bands between
0,15 MHz and 80 MHz (n)
80 % AM at 1 kHz
0% U
;
T
0°,45°,90°,135°,180°,225°,270°,315°
0% U
;
T
0% U
;
T
70%
0% U
;
T
0%
IEC 60601
Compliance level
Mains power quality should be that
± 2 kV
of a professional healthcare facility
100kHz Repetition
environment and Home healthcare
frequency
Mains power quality should be that
± 1 kV , Differential
of a professional healthcare facility
mode
environment and Home healthcare
± 2 kV Common mode
3 V
Mains power quality should be that
0,15 MHz – 80 MHz
of a professional healthcare facility
6 V m) in ISM and amateur
environment and Home healthcare
radio bands between
0,15 MHz and 80 MHz (n)
80 % AM at 1 kHz
Mains power quality should be that
of a professional healthcare facility
environment and Home healthcare
0,5 Cycles
1Cycle
25/30 Cycles (50/60Hz)
250/300 Cycles
(50/60Hz) (5s)
Electromagnetic environment -
guidance
environment.
environment.
environment.
environment.
.
If the user of the above listed
models requires continued
operation during power mains
interruptions, it is recommended
that the above listed models are
powered from an uninterruptible
power supply or battery.
ENG