7.2.1. Cleaning of the main unit (casing)
1. Use a cloth or sponge that is dampened with a mild detergent to clean the
external surfaces of the main unit.
2. Repeat the previous step with a mild disinfectant compatible with ABS and PC.
3. Use a cloth or sponge dampened with water and wipe the surfaces again. Dry
the surfaces using a clean cloth or a paper towel.
Never use dissolvent or abrasive products.
7.2.2. Decontamination of the liquids collection bottle
It is the user's responsibility to qualify any deviations from the recommended
method of processing.
Follow the decontamination product's instructions. Check the
decontamination product compatibility with the materials
After decontamination, label products with the date
The disposable jars should be processed according to the local rules.
After each use, decontaminate the collector recipient pursuant the next procedure:
A. Disconnect the 3 tubes and the filter and replace them for new ones.
B. Disassemble the cover and the floating device from the transparent jar:
Free the two locking crabs to open the jar, lift the cover and pull down the floating
device.
If dirt is observed in the rubber quad-rings of the floating device or in the locking
crabs, they can be disassembled (it is recommended to warm up in water at 50ºC
to expand the rubber rings).
C. Washing and rinsing (manual or mechanical):
Wet in lukewarm water (<43ºC) before adding any detergent; Wash parts
thoroughly in warm water using a compatible detergent (rub with brushes, if
necessary); Rinse thoroughly with sterile water, by immersing it completely in
three separate copious volumes of water. Each rinse should be a minimum of one
minute. Do not reuse the water. Finally, allow to dry.
• Alkaline or neutral detergents: non-ionic tension-actives concentration lower
than 15%, or ionic tension-actives concentration lower than 30%.
• Enzymatic detergents: enzymes concentration lower than 15%.
ENGLISH
9
Locking crabs
Floating device