General Specifications
General Specifications
General Specifications
General Specifications
Display: 3 1/2 digit LCD, 1999 cts
Polarity Indication: Automatic,
positive implied, negative indicated
Zero Adjustment: Automatic
Overrange Indication: (1.) or (-1.) is
displayed ("OL", 16XL)
Low Battery Indication: .
Change battery immediately.
Display Update Rate: 3/sec. ,
nominal; 2.5/sec. , nominal 16XL
Operating Temperature: 0°C to 50°C,
0 to 70% Relative Humidity
Storage Temperature: -20°C to 60°C,
0 to 80% RH, battery removed
Environment: Indoor use, Altitude up to
2000 m.
Power: 9-volt transistor battery, NEDA
1604, JIS 006P, IEC6F22
Battery Life (Typical): 300 hours
alkaline, 200 hours carbon-zinc
Dimensions (H x W x D): 6.1 x 2.8 x
1.3 inches, (15.5 x 7.2 x 3.2 cm)
Weight (incl. battery): 8 oz (210 gr)
Accessories: Test leads, spare fuse,
battery, and Operator's Manual.
Safety: (5/10/15XL)meets EN61010-1; Cat
Safety
Safety
Safety
II -1000VDC / 750VAC;(16XL) Cat III-
600V; Pollution degree 2; Class II.; EMC
Meets EN50081-1, EN50082-1
This product complies with requirements of the
following European
Community Directives:
89/336/EEC (Electro

SPECIFICATIONS

SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
+ -
EMC:
EMC
EMC
-28-
magnetic Compatibility) and 73/23/EEC
(Low Voltage) as amended by 93/68/EEC (CE
Marking). However, electrical noise or intense
electromagnetic fields in the vicinity of the
equipment may disturb the measurement
circuit. Measuring instruments will also
respond to unwanted signals that may be
present within the measurement circuit. Users
should exercise care and take appropriate
precautions to avoid misleading results when
making measurements in the presence of
electronic interference.
Electrical Specifications
Electrical Specifications
Electrical Specifications
Electrical Specifications
Accuracies at 23°C ± 5°C, < 75 % RH.
DC Volts
DC Volts
DC Volts
DC Volts
Ranges: 5/10/15XL: 200mV, 2, 20, 200,
1000V; 16XL: 200mV, 2, 20, 200, 600V
Accuracy, all Ranges
5XL: ±(0.8% rdg +1 dgt)
10XL: ±(0.7% rdg +1 dgt)
15/16XL: ±(0.5% rdg +1 dgt)
Resolution, 200mV Range: 100µV
Input Impedance: 10MΩ(5XL: 1MΩ)
OL Protection, 200mV Range:
350VAC RMS/500VDC (15 sec.)
All other Ranges
5XL: 600V RMS/1000VDC (60 sec.)
10XL & 15XL: 750VAC RMS /
1000VDC 16XL: 600VDC or AC RMS
AC Volts (50-500Hz). Accuracies
AC Volts
AC Volts
AC Volts
apply to sine wave inputs only
Ranges- 5XL: 200, 500V
10XL: 200, 750V
15XL: 200mV, 2, 20, 200, 750V
16XL: 200mV, 2, 20, 200, 600V

Hide quick links: