2.2.3 Environment Requirement
Item
Machine
Machine hall
floo r
Environm
ent
Relative
temperature
Pressure
Door & window
Wall
Fire protection
Power
2.2.3 Grounding Requirement
• All function modules' good grounding designs are the base of reliability and stability
of device. Also, they are the most important guarantee of lightning arresting
and interference rejection. Therefore, system must follow this rule.
• Coaxial cable's outer conductor and isolation layer should keep sound electric
conducting with the metal housing of device.
• Grounding conductor must adopt copper conductor in order to reduce
high frequency impedance, and the grounding wire must be as thick and
W hen user i n stall machi ne frame a r ra y in one
machine hall , the distance betwe en 2 ro w of ma
hal
chi ne frames should be 1.2 — 1.5 m
Electric Isolation, Dust Free
Volume resistively of ground anti-static material:1(107—
1(1010( Grounding current limiting resistance: 1M(
Floor bearing should be greater than 450Kg1m2
5—40(C sustainable ) 0—45(C short time)
installing air-conditioning is recommended
20%—80% sustainable
86—105KPa
Installing rubber strip for sealing door-gaps and
dual level glasses for window
It can be covered with wallpaper, or brightnessless paint.
Fire alarm system and extinguisher
Requiring device power, air-conditioning power and
lighting power are independent to each other. Device
power requires AC power 220V 50Hz. Please carefully
check before running.
Req u
irement
10%—90% short time