Klein Tools ET250 Manual De Instrucciones página 7

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

CONTINUITY
Remove power from circuit. Test for continuity by applying test leads
to the system being tested. The tester automatically enters continuity
testing mode. If resistance < 270kΩ is detected, an audible signal will
sound and the continuity icon " " will show in the display indicating
continuity. If the circuit is open, "
DO NOT attempt to measure continuity on a live circuit.
GFCI TESTING
CAUTION: Check the GFCI Receptacle user manual for more
information. Always contact a qualified electrician to resolve wiring
problems. Operate the Test Button on the GFCI Receptacle. If the
GFCI circuit does not trip, the receptacle is not operating properly.
CAUTION: To resolve wiring or GFCI concerns, contact a
qualified electrician.
To test the functionality of a receptacle protected by a Ground Fault
Circuit Interrupter (GFCI), apply leads to the hot/live and ground
terminals. Press the GFCI Test button on the ET250 tester.
• If the GFCI device is wired properly, the GFCI will trip and the
circuit will become de-energized. The ET250 tester will stop
indicating voltage.
• If the GFCI receptacle is incorrectly wired, the power to the circuit
will remain and the ET250 will continue to indicate voltage. Contact
a qualified electrician to resolve wiring problems.
NOTE: The GFCI Test mode on this tester is deactivated in the
presence of voltages > 135V.
OPERATING INSTRUCTIONS
" will show in the display.
7

Publicidad

Tabla de contenido
loading

Tabla de contenido