iCH Auto/English
3. Installation
3.1 Disassembling the Water Chamber
1. Turn the device off and allow the heater and water to cool. If necessary, disconnect the flexible
tubing from the water chamber.
2. Place thumb on the top of the water chamber and tip it upward as shown in Figure 3-1. Remove
the water chamber from the system.
3. Unlock the water chamber by pulling the tab as shown in Figure 3-2. Remove the chamber lid
from the base.
3.2 Filling the Water Chamber
1. Fill the chamber to the fill line
Figure 3-3.
2. If the silicon gasket has been removed, replace it securely on the top edge of the water
chamber base. Otherwise, the device will not deliver air properly.
3. Position the water chamber on the heater plate and push the bottom forward to lock it in place,
as shown in Figure 3-4. Make sure the silicon connector on the inlet connector fits securely over
the CPAP device air outlet.
CAUTION: Do not overfill the water chamber. Damage to the device may occur.
CAUTION: Use only room temperature water; do not fill the chamber with hot or chilled water.
CAUTION: Do not turn the heater on without the water chamber installed
CAUTION: When installing the water chamber, do not allow water to spill into the device.
CAUTION: Do not move the device when the chamber has water in it.
Figure 3-1
Figure 3-3
Figure 3-2
with distilled water (approx. 300 ml), as shown in
Figure 3-4
6