Performance
Flow rate
1
Positive Pressure
Safety Relief Valve:
Gauge Accuracy
Standard
High
Pediatric
Neonatal
C.A.S.S.
Physical
Dimensions
Weight
Battery
Lithium
without fittings at full increase setting depending on supply vacuum and open air flow.
1
full scale deflection
2
full range at 22°C
3

ENVIRONMENTAL SPECIFICATIONS

Operating Temperature Range:
Storage Temperature Range:
Operating and storage Relative Humidity:
Ingress Protection Ratings:

Specifications

Standard
High
80 LPM
80 LPM
Located in the vacuum supply line to prevent pressurization of
the patient connection by failed injector vacuum (Venturi) units, or
inadvertent cross connection to pressured gases.
Gauge Range
0-200 mmHg (0-26.7 KPA)
0-760 mmHg (0-101.3 KPA)
0-160 mmHg (0-21.3 KPA)
0-160 mmHg (0-21.3 KPA)
0-100 mmHg (0-13.3 KPA)
0-160 mmHg (0-21.3 KPA)
6.7"H x 2.8"W x 4.8"D
(17.02 cm x 7.1 cm x 12.2 cm)
1 lb 4 oz (0.57 kg)
Two 2/3 AA batteries, 3.6V, 1.6 Ah, lithium
50 to 104ºF (10 to 40ºC)
-13 to 158ºF (-25 to 70ºC)
5 to 95% RH (Non-condensing)
IP 20
8700-0007-000 (Rev. 14) 03/2020
Pediatric
Neonatal
80 LPM
80 LPM
Analog
2
±5% Full-scale
±5% Full-scale
±5% Full-scale
±5% Full-scale
±5% Full-scale
±5% Full-scale
C.A.S.S.
80 LPM
Digital
3
±1% Full-scale
±1% Full-scale
±1% Full-scale
±1% Full-scale
±1% Full-scale
±1% Full-scale
5