3.6 Accumulate Mode
Use this mode to store the weight and HI, LO, SUM statistics from a series of
samples.
• Press ZERO to set the display to zero.
• Place the item on the pan or tared container.
• Press M+ to store the weight of the item. Storage is confirmed by the display
[n----x], where x is the number of samples. The SUM LED will illuminate to
indicate there are readings in memory.
• Remove the item, wait for stable zero reading, then place the next item on the
pan. Press M+ to store the weight of the next item. Repeat this process until all
items have been weighed.
Review the accumulation data by pressing M+ when the display is at zero.
• The display shows the number of samples [n----x]
• Press M+ to view the maximum sample weight [Hxxxxx].
• Press M+ to view the minimum sample weight [Lxxxxx].
• Press M+ to view the sum of weights [xxxxx].
• Press M+ to return to weighing.
Clear the accumulated data by pressing ZERO when the number of samples [n----x]
is displayed.
Note:
• Only a total of 999 accumulated records [n 999] can be stored, and the sum
of the weights cannot exceed [999999]. [n OVEr] or [A OVEr] will be displayed
if these values are exceeded. Clear the Accumulate data to continue.
• Accumulated data is cleared when the unit of measure is changed.
3.7 Checkweigh Mode
Use this mode to compare samples to a target weight range (see section 4.1 to
activate and set limits).
Press ZERO to set the display to zero. Place a sample on the pan or tared container,
and read the displayed weight and under/accept/over status, where:
Under
= weight is below preset minimum value
Accept = weight is equal to or within preset target range
Over
= weight exceeds preset maximum value
Note: When the unit of measure is changed (see Section 4.2), checkweigh limits
and setting (see Section 4.1) will be reset to "0".
EN-5