3.4 Insertion of the Viscosity Measuring Unit ViscoClock into the thermostat bath
The transparent thermostat has to be equipped with a VZ 5402 manual measuring insert in order to ac-
commodate the Viscosity Measuring Unit ViscoClock.
• The Viscosity Measuring Unit ViscoClock, with filled viscometer, is inserted into the manual measuring
insert {22) of the thermostat bath.
Fig. 5
3.5 Connecting the Viscosity Measuring Unit ViscoClock
CAUTION
Prior to plugging the power supply (29) into the socket, make sure that the mains-unit voltage matches the
mains voltage.
• Plug the 9 V/ DC jack plug (19) into the back panel of the electronic measuring unit
• Plug the TZ 1858 universal voltage power supply for 100 - 240 V~, 50/60Hz, into the mains socket.
3.6 Connecting the Viscosity Measuring Unit ViscoClock to a computer (PC) or
printer
There is an RS-232-C Interface (18) on the rear side of the electronic measuring unit (13).
• lf desired, connect a computer (PC) or printer for documentation purposes.
Plug connection: 4-channel mini DIN round plug.
Required data cable, which is not included in the scope of delivery, → accessories, chapter 5.
Fig. 6
The transfer parameters of the RS-232-C Interface
are firmly set:
• - 4800 baud
• - 7 bit word length
• - 2 stop bits
• - no parity
Initial operation
Thermostat bath
surface